Develosil HSR AQ C18 and HSR C1 Specifications

Applications

Analysis of Water Soluble Vitamins

HPLC Analytical Conditions:
Column:  1. Develosil HSR AQ C18, 5 µm (4.6 x 150 mm)

2. Develosil HSR C1, 5 µm (4.6 x 150 mm)

Mobile Phase: Methanol/25mM NH₄H₂PO₄ (pH=2.0)=25/75
Gradient: B) 5%(0-2min), 5%-50%(2-7min), 50%(7-10min), 5%(10.1min-)​
Flow Rate: 1.0 mL/min
Temperature: 40°C
Detection: UV at 280 nm
Sample: 1. (-)-Epigallocatechin (0.25mg/mL)
2. (-)-Epigallocatechin gallate (0.30mg/mL)
3. Caffeine
4. (-)-Epicatechin (0.25mg/mL)
5. (-)-Epicatechin gallate (0.30mg/mL)
Injection Volume: 2.0 µL

Analysis of Catechin in Green Tea

HPLC Analytical Conditions:
Column:  1. Develosil HSR AQ C18, 5 µm (4.6 x 150 mm)

2. Develosil HSR C1, 5 µm (4.6 x 150 mm)

Mobile Phase: Methanol/25mM NH₄H₂PO₄ (pH=2.0)=25/75
Gradient: B) 5%(0-2min), 5%-50%(2-7min), 50%(7-10min), 5%(10.1min-)​
Flow Rate: 1.0 mL/min
Temperature: 40°C
Detection: UV at 280 nm
Sample: 1. (-)-Epigallocatechin (0.25mg/mL)
2. (-)-Epigallocatechin gallate (0.30mg/mL)
3. Caffeine
4. (-)-Epicatechin (0.25mg/mL)
5. (-)-Epicatechin gallate (0.30mg/mL)
Injection Volume: 2.0 µL

Analysis of Peptides

HPLC Analytical Conditions:

Column:  1. Develosil HSR C1, 5 µm (4.6 x 150 mm)
2. Develosil HSR AQ C18, 5 µm (4.6 x 150 mm)
Mobile Phase: A. 25 mM NaCl (pH 3.0)
B) Acetonitrile
Flow Rate: 1.0 mL/min
Gradient: See more data here.
Temperature 30°C
Sample: 1.Gly-Tyr (0.05mg/mL)
2.Val-Tyr-Val (0.05mg/mL)
3.Met-Enkephalin (0.05mg/mL)
4.Angiotensin II (0.05mg/mL)
5.Leu-Enkephalin (0.05mg/mL)
Injection Volume:  10.0 µL